Abstract
Abstract. Although dialyzer reuse in chronic hemodialysis patients is commonly practiced in the United States, performance of reused dialyzers has not been extensively and critically evaluated. The present study analyzes data extracted from a multicenter clinical trial (the HEMO Study) and examines the effect of reuse on urea and β2-microglobulin (β2M) clearance by low-flux and high-flux dialyzers reprocessed with various germicides. The dialyzers evaluated contained either modified cellulosic or polysulfone membranes, whereas the germicides examined included peroxyacetic acid/acetic acid/hydrogen peroxide combination (Renalin®), bleach in conjunction with formaldehyde, glutaraldehyde or Renalin, and heated citric acid. Clearance of β2M decreased, remained unchanged, or increased substantially with reuse, depending on both the membrane material and the reprocessing technique. In contrast, urea clearance decreased only slightly (approximately 1 to 2% per 10 reuses), albeit statistically significantly with reuse, regardless of the porosity of the membrane and reprocessing method. Inasmuch as patient survival in the chronic hemodialysis population is influenced by clearances of small solutes and middle molecules, precise knowledge of the membrane material and reprocessing technique is important for the prescription of hemodialysis in centers practicing reuse.
High-flux hemodialyzers and reuse of dialyzers have been widely used for decades, yet the effects of these practices on solute clearances have not been fully evaluated. Although the beneficial effect of increasing urea clearance on clinical outcome has been established, at least up to a single-pool Kt/V value of 1.2 (1), there are also accumulating data suggesting that the removal of middle molecules (using vitamin B12 as marker) influences patient survival (2,3). Thus, maintenance of the clearance of both small and large solutes for reused dialyzers is important.
Reuse can affect dialyzer performance in at least two different ways. The first is the result of deposition of blood elements inside the lumen of the blood compartment and onto the dialyzer membrane. The second is the result of the reprocessing procedure. At present, the popular germicides used in reprocessing in the United States are Renalin (made up of peroxyacetic acid, acetic acid, and hydrogen peroxide, Minntech, Minneapolis, MN), formaldehyde, and glutaraldehyde (4). To enhance the aesthetic appearance of the dialyzers during reuse, sodium hypochlorite (bleach) is often used in conjunction with formaldehyde or glutaraldehyde to remove residual blood proteins. More recently, heated citric acid has also been introduced to clean and disinfect dialyzers for reuse. Because the chemical composition and mechanical structure are vastly different among various types of dialysis membranes, their interactions with the blood elements and reprocessing agents are likely to differ as well.
The HEMO Study is a prospective randomized multicenter trial sponsored by the U.S. National Institutes of Health designed to examine the effects of urea Kt/V and the type of dialysis membrane on clinical outcome of chronic hemodialysis patients (5). Various models of dialyzers and reprocessing methods are used among the 15 clinical centers (more than 45 dialysis units) in the trial. Using this large database, we have prospectively examined the effects of various combinations of dialyzers and reprocessing agents on the clearance of urea and β2-microglobulin (β2M). The data show that the effects of reuse on β2M are far more drastic than those on urea clearance. Furthermore, the effects vary greatly depending on the dialysis membrane material and reprocessing reagents. These observations confirm and extend our fundamental understanding of alterations in dialyzer performance during reuse.
Materials and Methods
Dialysis Prescriptions
The basic design of the HEMO Study has been described (5); therefore, only the parts of the protocol that are germane to the current objectives will be reiterated. Among other criteria, patients must have residual renal clearance of urea below 1.5 ml/min per 35 L of urea distribution volume to qualify for the study. Eligible patients were randomized to one of four groups: (1) standard urea Kt/V and lowflux membrane; (2) standard urea Kt/V and high-flux membrane; (3) high urea Kt/V and low-flux membrane; and (4) high urea Kt/V and high-flux membrane. The two standard urea Kt/V groups had a dialysis treatment target equilibrated Kt/V (eKt/V) of 1.05, whereas the two high urea Kt/V groups have a target eKt/V of 1.45. eKt/V is a double-pool value that accounts for the postdialysis rebound of plasma urea concentration and is approximately 0.2 lower than the single-pool value under usual clinical conditions in the United States (6). Blood flow rates, dialysate flow rates, models of dialyzer, and treatment durations were tailored to individual patients to achieve the target eKt/V. Although the nephrologists were encouraged to achieve the target eKt/V within the shortest duration possible, the minimum acceptable time was 2.5 h per session and varied depending on the necessity to remove the intradialytic fluid gain from individual patients.
The model of dialyzer used for each patient was chosen by their respective dialysis unit medical director and/or primary nephrologist as long as it satisfied the transport characteristics for urea and β2M as described below. Therefore, large spectra of blood flow rates, dialysate flow rates, dialyzer models, and treatment durations were encountered in the study.
Hemodialyzers
Urea KoA values for all dialyzer models were tested in vitro in a single laboratory at the University of Utah, and the results of these tests have been reported previously (7). Each dialyzer used in the HEMO Study had an in vitro urea KoA value ≥500 ml/min at a dialysate flow rate of 500 ml/min. Before acceptance into the HEMO Study, all dialyzer models were also tested for β2M clearance on at least five separate dialyzers during routine clinical hemodialysis in the HEMO Pilot Study or in special studies arranged by the HEMO Study. A low-flux dialyzer was defined as one that had a mean clearance of β2M < 10 ml/min during first use. In contrast, a high-flux dialyzer was defined as one that satisfied both of the following criteria: (1) an ultrafiltration coefficient (Kuf) > 14 ml/h per mmHg and (2) a mean β2M clearance > 20 ml/min during first use or over the lifetime of the dialyzer model with a given reprocessing technique.
Baxter Corp. (McGaw Park, IL) and Fresenius (Walnut Creek, CA) donated many of the dialyzers used in the HEMO Study. The most common low-flux dialyzers used were the CA series (CA170 and CA210, cellulose acetate membrane, Baxter) and F8 (polysulfone membrane, Fresenius). The most common high-flux dialyzers were CT190 (cellulose triacetate membrane made by Toyoba and marketed as dialyzers by Baxter) and the F80 series (F80A and F80B, polysulfone membranes, Fresenius). The F80A membrane is more porous and is designed for no reuse or reuse without bleach, whereas F80B is less porous and is designed for reuse with bleach because bleach has previously been shown to induce albumin loss through porous high-flux polysulfone membranes (8,9,10). Although other dialyzers were also used by some centers, either they were not reused or their sample sizes were relatively small. Therefore, analyses were confined to the most commonly used dialyzers in this report. The performance of F80 dialyzers provided by Fresenius changed in early 1997. The new F80A and F80B dialyzers appeared to be more porous since β2M clearances during first use were greater than those of their earlier counterparts (see below).
Reuse Processing Techniques
The reprocessing techniques used in the HEMO Study are listed in Table 1. The technique used by a particular dialysis unit was determined by the policy of the unit and was not altered because of its participation in the HEMO Study. Because the dialyzer models were also chosen by the individual clinical centers, a particular reprocessing method might only be applied to some dialyzer models and not others. For example, the two centers that reprocessed dialyzers using heated citric acid only used polysulfone dialyzers from Fresenius. No other type of membrane so far has been shown to be suitable for heat reprocessing. Regardless of the reprocessing technique, a residual total cell volume of ≥80% of the original value was a prerequisite for reuse of the dialyzer.
Reprocessing techniques used in the study
Restrictions on Number of Reuses
The maximum number of times that a dialyzer could be reused in the HEMO Study was originally set at 20. Individual dialysis units could also restrict reuse to less than 20 times. For example, the two centers that used heated citric acid limited their dialyzers to 12 and 15 reuses, respectively. Monitoring of data during the course of the study revealed a significant reduction in the β2M clearance of certain high-flux dialyzers when they were reprocessed with Renalin without bleach (see below). To maintain a clear separation in β2M clearances between the low-flux arm and the high-flux arm of the study, the maximum allowable number of reuses for dialysis units using Renalin without bleach was reduced to 10 in February 1997, and reduced further to 6 in October 1997. In dialysis units where Renalin was used in conjunction with bleach, the maximum allowed number of reuses remained at 20.
Data Collection
Data for this report were collected after randomization of the patients into the follow-up phase of the HEMO Study. Only data from March 1995 (initial enrollment of patients into the full-scale phase of the HEMO Study) through February 1998 were included. Before June 1997, predialysis and postdialysis β2M concentrations were obtained for determination of β2M clearances during kinetic modeling sessions at months 1, 2, 3, 4, and 6 after randomization into the treatment groups, and every 6 mo thereafter. Beginning in June 1997, the frequency of the β2M measurements was increased to months 1, 2, and every 2 mo thereafter in the high-flux arm (both high and standard urea eKt/V) to allow more rapid understanding of dialyzer performance with various reprocessing techniques. At the same time, the frequency of β2M measurements in the low-flux arm (both standard and high urea eKt/V) was reduced to months 1, 12, and every 12 mo thereafter.
This report includes all available β2M clearance determinations for the specified dialyzers. Analyses of β2m clearance for the high-flux F80 dialyzers were separated into those obtained before June 15, 1997, and those obtained thereafter (June 15 through February 1998) because of the changes in the dialyzers by the manufacturer. The numbers of β2M clearance measurements used in the analyses of specific dialyzer-reuse combinations are provided in Table 3.
Reuse-averaged β2M clearances of high-flux dialyzers through different numbers of reuses with various reprocessing methods
Predialysis and postdialysis blood urea nitrogen (BUN) concentrations were measured at monthly kinetic modeling sessions. In this report, analyses of urea clearance were restricted to these modeling sessions performed in patients with functional arteriovenous (native or polytetrafluoroethylene) vascular accesses. Exclusions from these analyses of urea clearances were patients with access recirculation >15% determined using the slow-flow method (assessed once before randomization and once at the 4-mo follow-up visit), patients using temporary catheters, patients with extremity amputation(s), sessions with more than 15 min of interruption (which accounted for 2.3% of modeling sessions), and sessions with highly deviant predialysis or postdialysis BUN concentrations (postdialysis BUN <10% or >60% of the predialysis value, or predialysis BUN >200 mg/dl or <10 mg/dl, accounting for a total of 1.9% of modeling sessions).
Predialysis and postdialysis body weights, location of the vascular access, dialyzer model, dialyzer blood flow rate, type of delivery system, dialysate flow rate, ultrafiltration rate, and total ultrafiltration volume were noted or calculated for each session. Blood samples were collected immediately predialysis and 20 s postdialysis after the dialyzer blood flow rate had been reduced to ≤80 ml/min.
Assays
Blood samples were immediately centrifuged and all serum and plasma samples were assayed for urea nitrogen and β2M concentrations at a central laboratory (LifeChem, Rockleigh, NJ). Urea was measured by an autoanalyzer (Hitachi 747 to 200, Boehringer Mannheim, Indianapolis, IN) using a kinetic urease method. The intra-assay and interassay variations were 1.7 and 3.0%, respectively. β2M concentrations were determined using a solid-phase competitive RIA with reagents supplied by Abbott Laboratories (Abbott Park, IL) and radioactivity determined by a Micromedic Apex Automatic Counter (model 10600; ICN Biomedicals, Costa Mesa, CA). The intra-assay and interassay variations were 3.6 and 5.0%, respectively.
Calculation of β2M Clearance
Dialyzer clearance of β2M (Kd) was calculated using the equation recently described (11): where C(0) and C(T) denote the predialysis and postdialysis concentrations of β2M, respectively; Qf denotes the average ultrafiltration rate throughout the entire treatment session of length T and is calculated from predialysis and postdialysis body weights; and V denotes the distribution volume for β2M. In the calculation, several assumptions were made: (1) The distribution volume of β2M = extracellular volume = 1/3 × urea distribution volume determined by kinetic modeling. (2) Fluid removed during hemodialysis was derived exclusively from the extracellular compartment. (3) Intradialytic generation and patient residual renal and nonrenal clearance of β2M were negligible. (4) Postdialysis rebound of plasma β2M concentration was ignored. Clearance estimated by this equation yields an average value over the entire treatment and includes removal by diffusive, convective, and absorptive mechanisms.
Calculation of Effective and Predicted Urea Clearance
The patient's predicted urea clearance was estimated from the in vitro mass transfer area coefficient (KoA), blood flow rate, and dialysate flow rate using a standard formula (7), with appropriate adjustments for blood water concentration of urea (12,13) and ultrafiltration rate. The KoA values used in these calculations were determined in vitro for individual dialyzer models in a single laboratory as reported previously (7) and described above.
To determine effective urea clearance, the patient's anthropometric volume (14) was adjusted for two pool effects to obtain an estimate of the patient's single-pool volume using previously described methods (15). The adjusted anthropometric volume was then used as an input parameter in a two-BUN method (16), along with the predialysis and postdialysis BUN values to compute the effective clearance.
The ratio of the effective to the predicted clearance was calculated and used as the dependent variable in multivariate regression analyses to assess the effect of reuse on effective urea clearance.
Statistical Analyses
All data are presented as mean ± SEM unless specified otherwise. The relationship between β2M clearance and reuse number was described for specific dialyzer-reuse combinations by box-plots (17), and by regression of β2M clearance on reuse number using a mixed effects model with a compound symmetry error structure to account for multiple measurements in the same patients (18). For each dialyzer-reuse combination, linearity of the relationship between β2M clearance and reuse number was assessed by testing the significance of a quadradic term in the regressions. Mixed-effects regression models were also used to test for the presence of a relationship between ultrafiltration rate or total ultrafiltration volume in the session and reuse number within each dialyzer-reuse combination.
The time-weighted mean β2M clearance over a given number (k) of reuse was estimated for each dialyzer-reuse combination by first computing the mean β2M clearance for that dialyzer-reuse combination at each reuse number, and then averaging the values of all uses up to reuse number k. To relate effective urea clearance to reuse number, the ratio of the effective to the predicted urea clearance was regressed on reuse number while controlling for the vascular access location, type of dialyzer, dialyzer blood flow rate, type of delivery system, dialysate flow rate, and race of the patient in a multivariate regression analysis. This regression was also carried out using a mixed-effects model with a compound symmetry error structure.
Results
Characteristics of Patients and Hemodialysis Sessions
The characteristics of the studied patients and treatment sessions are listed in Table 2. The data are presented separately for the low-flux and high-flux groups because the effect of reuse on β2M clearance was far more substantial than that on urea clearance (see below). Body weights, volumes of distribution of urea, blood flow rates, dialysate flow rates, treatment session durations, and ultrafiltration rates were all similar between the high-flux and low-flux groups. By design, mean ± SD β2M clearance over all reuse sessions for patients in the high-flux group (34.6 ± 14.3 ml/min) was approximately 11-fold higher than that in the low-flux group (3.1 ± 8.1 ml/min).
Characteristics of patients and hemodialysis sessionsa
β2M Clearance by High-Flux and Low-Flux Dialyzers during First Use
There was no substantial difference between β2M clearance during first use by dialyzers that had been processed and those that had not been processed before use (data not shown). Mean clearances of β2M for all low-flux dialyzer models during first use were well below 10 ml/min. For example, mean ± SD β2M clearance by CA210 (n = 309) and F8 (n = 279) were 3.1 ± 11.0 and 1.7 ± 11.3 ml/min, respectively. Mean clearances of β2M by high-flux dialyzers during first use were more variable, ranging from 14.2 ± 13.4 ml/min (n = 66) for F80B that were used early in the study to 42.3 ± 12.5 ml/min (n = 336) for CT190 dialyzers (Figure 1). The F80A and F80B dialyzers provided by the manufacturer changed during early 1997 and mean β2M clearances by these dialyzers during first use have increased. It should be noted that the variations of β2M clearances by individual dialyzers within any given low-flux or high-flux model were large, with SD ranging from 9.3 to 16.7 ml/min.
β2-microglobulin (β2M) clearance during first use by CT190 (cellulose triacetate high-flux), F80A, and F80B (polysulfone high-flux) dialyzers. The performance of F80 dialyzers changed during early 1997. Mean β2M clearances were significantly higher after June 15, 1997 than those observed before this date for the F80B (P = 0.006) and F80A (P = 0.002) dialyzers, whereas no significant changes were observed between these two periods for the CT190 dialyzers (P = 0.15). The middle horizontal line inside each box represents the median; the plus sign (+) represents the mean; the lower and upper horizontal lines marking the bottom and top of the box represent the 25th and 75th percentiles, respectively; the lower and upper vertical lines extend to the 10th and 90th percentiles, respectively; and the hatched region represents the 95% confidence interval for the mean.
Effect of Renalin without Bleach on β2M Clearance
Reuse using Renalin as germicide had a great effect on β2M clearance by CT190 dialyzers. As can be seen in Figure 2, mean β2M clearance declined rapidly through the first four reuses and continued to decline at a slower rate up to the tenth through 14th reuses. The mean ± SEM total decrease in β2M clearance between first use and the tenth through 14th reuses was 67.2 ± 2.7%. The clearance did not appear to decrease further when reuse was extended to 15 to 20 times. Since the β2M clearance during a given session decreased with increasing reuse number, so did the time-averaged clearance through the life of the dialyzers (Table 3). Similar results were observed in all six clinical centers that used this dialyzer-reprocessing combination (data not shown).
Effect of reprocessing using Renalin on β2M clearance by CT190 (cellulose triacetate high-flux) dialyzers. The data were obtained from the full study period. The number of observations for each range of reuse numbers is presented above the x-axis. The mean ± SEM β2M clearance declined by 67.2 ± 2.7% between first use and 10th to 14th reuses. For explanation of box-plots, see the legend to Figure 1.
Reprocessing with Renalin produced a slight decline in β2M clearance by the F80A dialyzers (the change was -0.64 ± 0.26 ml/min per reuse; P = 0.013 before June 15, 1997, and -0.31 ± 0.31 ml/min per reuse; P = 0.32 after June 15, 1997) (Figure 3).
Effect of reprocessing using Renalin on β2M clearance by F80A (polysulfone high-flux) dialyzers. The data were obtained since June 15, 1997. The mean ± SEM of the intercept of the regression of β2M clearance versus reuse numbers is 37.9 ± 1.5 ml/min, and the slope is -0.31 ± 0.31 ml/min per reuse (P = 0.32). For explanation of box-plots, see the legend to Figure 1.
Effect of Bleach on β2M Clearance
Bleach was always used in conjunction with formaldehyde, Renalin, or Diacide® for reprocessing of dialyzers in the HEMO Study.
F80B Dialyzers. For F80B dialyzers, the rate of increase of β2M clearance with reuse differed significantly (P < 0.001) among the three germicides (formaldehyde, Renalin, and Diacide) used in conjunction with bleach. Bleach in conjunction with formaldehyde had the greatest effect on F80B dialyzers. β2M clearance by F80B dialyzers increased markedly through 20 reuses (Figure 4A). Accordingly, the time-averaged clearance increased by more than twofold through 20 reuses (Table 3). The mean ± SEM rate of increase was greater during the initial seven reuses (3.32 ± 0.40 ml/min per reuse; P < 0.001) than during the seventh through 20th reuses (1.40 ± 0.26 ml/min per reuse; P < 0.001).
(A) Effect of reprocessing using formaldehyde with bleach on β2M clearance by F80B (polysulfone high-flux) dialyzers. The data were obtained since June 15, 1997. The regression of β2M clearance versus reuse numbers is significantly nonlinear (P = 0.002). The intercept is 24.3 ± 2.0 ml/min. The slope for 0 to 7th reuses is 3.32 ± 0.40 ml/min per reuse (P < 0.001), whereas the slope for 7th to 20th reuse is lower at 1.40 ± 0.26 ml/min per reuse (P < 0.001). The mean ± SEM β2M clearance increased by 150 ± 39% between first use and 15th to 20th reuses. (B) Effect of reprocessing using Renalin with bleach on β2M clearance by F80B (polysulfone high-flux) dialyzers. The data were obtained since June 15, 1997. The mean ± SEM of the intercept of the regression of β2M clearance versus reuse number is 24.1 ± 2.6 ml/min, and the slope is 1.88 ± 0.26 ml/min per reuse (P < 0.001). (C) Effect of reprocessing using Diacide (glutaraldehyde) with bleach on β2M clearance by F80B (polysulfone high-flux) dialyzers. The data were obtained since June 1997. The mean ± SEM intercept of the regression of β2M clearance versus reuse number is 18.9 ± 1.6 ml/min, and the slope is 0.55 ± 0.18 ml/min per reuse (P = 0.004). For explanation of box-plots, see the legend to Figure 1.
The effectiveness of bleach in increasing β2M clearance of F80B dialyzers was lower when it was used in conjunction with Renalin (1.88 ± 0.26 ml/min per reuse) (Figure 4B) instead of formaldehyde. When Diacide was used instead of formaldehyde or Renalin, the effect of bleach was even less (Figure 4C). β2M clearance of F80B dialyzers reprocessed using Diacide with bleach increased only by 0.84 ± 0.25 ml/min per reuse (P < 0.001) before June 15, 1997, and by 0.55 ± 0.18 ml/min per reuse (P = 0.004) after June 15, 1997.
CT190 Dialyzers. When bleach was used in conjunction with either formaldehyde or Diacide to reprocess CT190 dialyzers, modest increases in β2M clearance were observed (0.29 ± 0.11 ml/min per reuse; P = 0.011) (Figure 5). The rate of increase in β2M clearance did not differ significantly between the two aldehydes (0.32 ± 0.16 ml/min per reuse, P = 0.045 for formaldehyde with bleach, and 0.22 ± 0.17 ml/min per reuse, P = 0.20 for Diacide with bleach).
Effect of reprocessing using formaldehyde or Diacide (glutaraldehyde) with bleach on β2M clearance by CT190 (cellulose triacetate high-flux) dialyzers. The data were obtained from the full study period. The mean ± SEM of the intercept of the regression of β2M clearance versus reuse numbers is 40.9 ± 0.8 ml/min, and the slope is 0.29 ± 0.11 ml/min per reuse (P = 0.011). For explanation of box-plots, see the legend to Figure 1.
Low-Flux Dialyzers. When low-flux dialyzers were reprocessed using any procedure that included bleach, there was a small but statistically significant increase in β2M clearance during reuse (0.25 ± 0.07 ml/min per reuse; P < 0.001) (Figure 6). The large increase in β2M clearance observed with F80B dialyzers reprocessed with bleach and formaldehyde (see above) raised the possibility that low-flux dialyzers (F8) made from the same polymer (polysulfone) might exhibit similar behavior. When F8 dialyzers (0.19 ± 0.07 ml/min per reuse; P = 0.010) and low-flux dialyzers made from other materials (0.52 ± 0.18 ml/min per reuse; P = 0.004) were analyzed separately and compared, the rates of increase in β2M clearance with reuse tended to be greater for the nonpolysulfone dialyzers, but the difference was not statistically significant (P = 0.09).
Effect of reprocessing using formaldehyde, Diacide (glutaraldehyde), or Renalin with bleach on β2M clearance by all low-flux dialyzers. The data were obtained from the full study period. The mean ± SEM of the intercept of the regression of β2M clearance versus reuse numbers is 2.5 ± 0.6 ml/min, and the slope is 0.25 ± 0.07 ml/min per reuse (P < 0.001). For explanation of box-plots, see the legend to Figure 1.
The greater effect of formaldehyde than Diacide in conjunction with bleach on F80B dialyzers suggests that this difference between the two aldehydes might also be present in low-flux dialyzers. Further analysis showed that the rate of increase in β2M clearance with reuse of low-flux dialyzers did not differ significantly among formaldehyde, Diacide, and Renalin (all in conjunction with bleach) (P = 0.075), but tended to be greater with formaldehyde (0.33 ± 0.09 ml/min per reuse; P < 0.001) than with Diacide (-0.06 ± 0.15 ml/min per reuse; P = 0.71) or Renalin (0.13 ± 0.14; P = 0.37).
Effect of Heated Citric Acid on β2M Clearance
Polysulfone high-flux (F80A) dialyzers were also susceptible to the effect of heated citric acid. The clearance of β2M increased in an apparently linear manner up to seven reuses at a rate of 2.31 ± 0.45 ml/min per reuse (P < 0.001) (Figure 7). The effect of heated citric acid appeared to have reached a plateau phase after seven reuses. The mean increase in clearance was 36 ± 8% between first use and fifth through ninth reuses. The time-averaged β2M clearance increased accordingly with increasing reuse numbers (Table 3). In contrast, heated citric acid had no effect on β2M clearance by F8 dialyzers (0.00 ± 0.27 ml/min per reuse; P = 0.99).
Effect of reprocessing using heated citric acid on β2M clearance by F80A (polysulfone high-flux) dialyzers. The data were obtained since June 1997. The regression of β2M clearance versus reuse numbers is significantly nonlinear (P = 0.012). The mean ± SEM of the intercept is 43.6 ± 2.1 ml/min. The slope for 0 to 7th reuse is 2.31 ± 0.45 ml/min per reuse (P < 0.001), whereas the slope for 7th to 14th reuse is 0.01 ± 0.69 ml/min per reuse (P = 0.98). The mean β2M clearance increased by 36 ± 8% between first use and 5th to 9th reuse. For explanation of box-plots, see the legend to Figure 1.
Relationship between Reuse and Ultrafiltration Volumes
Transmembrane convective transport may contribute significantly to the dialyzer clearance of middle molecules, such as β2M. Since convective solute transport across a given membrane is dependent on ultrafiltration, the correlation between reuse numbers and ultrafiltration volumes was examined to determine whether the changes in β2M clearances during reuse were due to coincidental variabilities in ultrafiltration volumes. Regression analyses, however, showed no correlation between these two variables for any combination of high-flux or low-flux dialyzers and reprocessing methods (data not shown).
Effect of Reuse on Urea Clearances
The effects of reuse on urea clearances were, in general, more modest than those on β2M clearances, although both high-flux and low-flux dialyzers were affected (Table 4). Urea clearance by high-flux dialyzers decreased by 1.9 ± 0.3% per 10 reuses (P < 0.001) with no significant differences among various types of dialyzers and reprocessing methods studied (P = 0.096). For example, in contrast to their opposing effects on β2M clearances, Renalin (-2.9 ± 0.4% per 10 reuses for CT190 and -2.1 ± 1.0% per 10 reuses for F80A) and bleachcontaining methods (-1.6 ± 0.8% per 10 reuses for CT190 and -1.4 ± 0.5% per 10 reuses for F80B) had similar deleterious effects on urea clearances of high-flux dialyzers.
Effect of reuse on urea clearance
Urea clearances by low-flux dialyzers also deteriorated with reuse (-1.0 ± 0.3% per 10 reuses; P < 0.001), but the decrease was slightly lower than that of high-flux dialyzers (P = 0.015 for all low-flux dialyzers combined versus all high-flux dialyzers combined) (Table 4). There were no significant differences in decreases of urea clearances between Renalin (-1.1 ± 0.4% per 10 reuses) and bleach-containing methods (-1.2 ± 0.4% per 10 reuses) for low-flux dialyzers.
Discussion
Current Practice in the United States
The survey results from a 1995 study, which was recently published by the Center for Disease Control and Prevention, showed that 50% of the dialysis units in the United States treated some patients with high-flux dialyzers (4). High-flux dialyzers were defined in that report as ones that had ultrafiltration coefficients ≥20 ml/h per mmHg; β2M clearances were not part of the criteria. The report also showed that 77% of all units and 76% of all patients reused dialyzers. Renalin was the most commonly used germicide, accounting for 54% of all units reusing dialyzers. Formaldehyde and glutaraldehyde were used by 38 and 7% of the units, respectively, whereas 1.4% of the units that reused dialyzers disinfected with heat. It should be noted that peroxyacetic acid (a constituent of Renalin) in conjunction with bleach is not recommended for reprocessing dialyzers because the combination may produce hydrochloric acid vapors (19). Therefore, the bleach must be thoroughly rinsed off before the application of Renalin, or other germicides.
Effect of Reuse on Clinical Outcome
The effect of dialyzer reuse on patient mortality has been controversial, but two commonly cited studies have reported that reuse in the United States was associated with higher mortality rates (20,21). There are a number of reasons that could potentially explain these associations, some of which have been analyzed in more detail (e.g., free-standing versus hospital-based, type of germicides used), whereas others have not been well addressed (e.g., dose of dialysis). Dose of dialysis, as determined by Kt/V of urea, has been established to be a determinant of clinical outcome in chronic hemodialysis patients (1). Preliminary data suggest that middle molecule removal may also influence patient survival (2,3). To the extent that patient mortality may be affected by reuse of dialyzers, the effects of dialyzer reuse on small and middle molecule clearance should be better understood.
HEMO Study as a Database
At various time points of the HEMO Study, all participating centers have practiced dialyzer reuse, employing various reprocessing techniques (Table 1). In addition, the HEMO Study is, in essence, a trial of the importance of small and middle solute clearance, using urea and β2M as markers, respectively. The large number of enrolled patients, together with frequent assessments of urea and β2M kinetics according to strict protocols, provides a rich data source for the effect of reuse on dialyzer performance. The standard (1.05 target) and high (1.45 target) urea eKt/V groups also provide a wide spectrum of urea clearances for the examination of reuse effects.
There are limitations in these data. First, only a limited number of combinations of dialyzer and reprocessing techniques were studied. Second, blood flow rates, dialysate flow rates, and vascular access locations were not standardized, although these variables were accounted for in the statistical analyses of urea clearance. Patients with vascular access recirculation >15% were excluded for analyses of urea clearance in this report. Although data from patients with recirculation <15% were not analyzed further for the effect of recirculation on urea Kt/V or β2M clearance, there was no systematic bias in data collection during different reuse numbers in this large data set. Third, although the heparin dosage likely remained relatively stable over time for a given individual patient, heparin dosage was not standardized in the HEMO Study. Fourth, individual dialysis units in the HEMO Study were allowed to impose a maximum number of reuses for their dialyzers, as long as that number did not exceed the maximum number set by the HEMO Study for that particular dialyzer-reprocessing technique combination. For these reasons, reliable data on the maximum reusability of specific dialyzer models could not be obtained.
Effect of Reuse on Urea Clearance
Gotch reported approximately 20 years ago that urea clearances were maintained at more than 90% of their original values when reused dialyzers retained 80% of their original total cell volume (22). Since then, this observation has formed the basis of the performance criterion for accepting dialyzers for reuse. Although this criterion was initially established only for low-flux cellulosic membrane hollow-fiber dialyzers with relatively low blood flow rates, subsequent studies have suggested that it is applicable to modern conditions as well. For example, Canaud et al. found no decrease in the predialysis to postdialysis urea reduction ratio (URR) at the tenth use after reprocessing with peroxyacetic acid compared with first use of polyacrylonitrile membrane dialyzers (Filtral 16, Hospal) operating at blood flow rates of 300 to 400 ml/min. Their data, however, were limited to only two patients using 14 dialyzers each (23). Ouseph et al. showed that both high-flux cellulosic dialyzers (AM-UP-75WET, Asahi, Tokyo; n = 8) and high-flux polysulfone (F80B; n = 5) dialyzers maintained their urea Kt/V at 12th and 15th use, respectively, after reprocessing with Renalin, as long as total cell volume remained more than 80% of the original value (24).
In contrast, several other studies have suggested that urea clearances can deteriorate with dialyzer reuse. The marked (49%) decrease in urea clearances after 15 uses reported by Delmez et al. was attributed to nonuniform dialysate flow in reused dialyzers and appeared to be associated only with a specific lot of low-flux cuprammonium (Clirans TAF, Terumo, Tokyo, Japan) dialyzers (25). Garred et al. found that when 102 high-flux polysulfone dialyzers (F60 and HF80, Fresenius) were reprocessed with peroxyacetic acid, the mass transfer coefficient of urea decreased by approximately 3% after 15 uses even though the blood volume decreased by only 1% (26). Murthy et al. found that while urea clearance by low-flux cuprammonium dialyzers (T220L, Terumo) did not change, that by F80B dialyzers decreased by 8.3 to 9.6% after 20 reuses (27). The number of patients in that study was small (n = 6). Sherman et al. examined the effect of reuse on delivered urea Kt/V in the routine clinical practice setting (28). They found that mean Kt/V decreased from 1.10 at a mean reuse number of 3.8 to a mean Kt/V of 1.05 at a mean reuse number of 13.8. When the germicides were analyzed separately, reprocessing with formaldehyde was associated with a significant (5.5%) reduction, with no changes in mean Kt/V observed with either Renalin or glutaraldehyde. The number of patients in the two latter categories was small. Dialyzer models were not specified in that report. More recently, Leypoldt et al. studied the clearance of urea, creatinine, and phosphate by two low-flux (TAF175, Terumo and CA210) and two high-flux (CT190 and F80A) dialyzers reprocessed with Renalin (29). They showed a trend for the clearances of these solutes to decrease with reuse up to 14 times, although the changes were not statistically significant. The number of patients examined was limited to six.
Our present study confirmed that urea clearance indeed decreased with dialyzer reuse, although only to a modest extent. Furthermore, the decrease was observed with both high-flux and low-flux membranes, independent of the reprocessing technique. The large sample size in this study allowed for the detection of even relatively small changes. Reuse was limited to a maximum of 20 times (and less for some dialyzer-reprocessing technique combinations) in the HEMO Study. Although the magnitude of decrease in urea clearance was small (approximately 1 to 2% per 10 reuses, Table 4), the relationship between the decrease and reuse number appeared to be linear. If this relationship holds true beyond 20 reuses, the decrease in urea Kt/V could be substantial for dialyzers that are reused many times and may potentially contribute to the reported higher mortality associated with reuse.
Effect of Reuse on β2M Clearance
Previous studies of the effect of Renalin on β2M clearance by high-flux dialyzers produced conflicting results. Westhuyzen et al. reported no change in β2M removal by polyacrylonitrile (AN69 HF®) at fourth use, but that by polymethylmethacrylate (Toray) decreased by approximately 50% (30). Ouseph et al. showed that β2M clearance by high-flux cellulosic dialyzers (AM-UP-75WET, Asahi; n = 8) reprocessed with Renalin decreased at the 12th use, but the decrease was statistically insignificant. In contrast, dialytic reduction in the plasma β2M concentrations by high-flux polysulfone dialyzers (F80B; n = 5 to 9) reprocessed with Renalin decreased from 30% at first use to 12% at tenth use (P = 0.042) (24). Leypoldt et al. studied β2M clearance and total protein loss in the dialysate using CT190 and F80A dialyzers reprocessed with Renalin up to 14 times (29). Clearance of β2M decreased with reuse for both CT190 and F80A dialyzers. Mean dialysate total protein loss was less than 2 g per session for both types of dialyzers and also decreased with reuse.
In the present study, we found that reprocessing using Renalin without bleach substantially decreased the β2M clearance of CT190 dialyzers (Figure 2), with similar results observed in all six clinical centers using this technique. This is consistent with the hypothesis that adsorption of plasma protein onto the cellulose triacetate membrane surface forms a secondary membrane that diminishes the effective pore size of the dialysis membrane for β2M transport (29,31) and that Renalin alone is relatively ineffective in removing this protein layer. This hypothesis is also supported by the observation that complement activation is markedly attenuated when reused cellulosic membranes are reprocessed with peroxyacetic acid (32). Presumably, the deposited C3 fragments and other plasma proteins that are not removed by peroxyacetic acid inhibit further binding of C3 molecules for complement activation. Reprocessing using Renalin has only a small effect on the β2M clearance of F80A dialyzers (Figure 3), suggesting that either the polysulfone membrane is more resistant to fouling by plasma proteins such that its transport and/or adsorptive capacities are largely preserved, or that these protein—membrane interactions are more susceptible to disruption by Renalin.
Inclusion of bleach in the reprocessing procedure produces a very different effect. When bleach was used in conjunction with formaldehyde, β2M clearance by F80B dialyzers is enhanced substantially (Figure 4A), consistent with previous reports of increases in ultrafiltration coefficients (33), β2M clearance (9,27,33,34), and albumin losses (8,9,10) using the F80-bleach combination. The mechanism by which bleach produces these effects is unclear. The polysulfone dialysis membranes manufactured by Fresenius are made from copolymers of polysulfone and polyvinylpyrrolidone (PVP). It has been postulated that bleach removes PVP from the copolymers, thereby increasing the porosity of these membranes. It should be noted, however, that the evidence supporting this hypothesis was derived under conditions that were vastly different from those used in clinical practice. In those experiments, in vitro exposure of F60 dialyzers to 4% (compared with ≤1% clinically) of bleach for 5 d (compared with ≤5 min clinically) resulted in a loss of 38% in PVP content of the membrane (35).
When bleach was used in conjunction with Diacide instead of formaldehyde, the increases in β2M clearance by F80B dialyzers during reuse were markedly diminished (Figure 4C). The reason for this difference between the two aldehydes is also unclear. If the efficacy of bleach to increase membrane porosity is independent of formaldehyde, it would be reasonable to postulate that the glutaraldehyde applied to the dialyzers after bleach during the reprocessing procedure somehow reverses the effect of bleach. Both formaldehyde and glutaral-dehyde cross-link proteins. In contrast to formaldehyde, however, the reaction produced by glutaraldehyde is rapid and irreversible (36). Theoretically, glutaraldehyde would be more effective in cross-linking any residual plasma proteins and peptides that are left after the bleach cycle. Aggregation of the cross-linked proteins in the microstructure of the membrane would therefore decrease its effective pore sizes. An alternative but less likely possibility is that glutaraldehyde may, in fact, cross-link plasma proteins to the dialysis membrane or between the membrane polymers themselves. Experimental evidence for this postulate is unavailable.
In contrast, β2M clearance by CT190 dialyzers only increased mildly after reprocessing with bleach (Figure 5), possibly because cellulose triacetate membranes are relatively resistant to oxidative damage by hypochlorite. Clearance of β2M by low-flux polysulfone dialyzers (F8) did not change after exposure to bleach, possibly because the size of the membrane pores was initially too small, such that even the damage by bleach did not enlarge the pores sufficiently to enhance the passage of small proteins. The ineffectiveness of low-flux polysulfone dialyzers to remove β2M also suggests that the mechanism of removal of this protein by polysulfone membranes is primarily transport across the membrane rather than adsorption onto the membrane surface.
Previous data regarding the effect of heated citric acid on dialyzer clearance of proteins are very limited. One previous study showed that in vitro clearance of cytochrome C (a protein with molecular weight similar to that of β2M) by F80 dialyzers reprocessed with heated citric acid remained unchanged after 11 reuses (37). In contrast, β2M clearance by F80A dialyzers in the present study increased by approximately 36% after the dialyzers had been reprocessed with heated citric acid for five to nine times (Figure 7). This difference in results between the previous and present studies is potentially due to differences in the solute marker used (cytochrome C versus β2M), the testing condition (in vitro versus clinical), sample size, or changes in the intrinsic properties of F80 dialyzers.
Increases in β2M clearance during reuse of high-flux polysulfone membranes reprocessed with either bleach or heated citric acid could potentially be associated with increases in albumin loss through the membrane, as reported with polysulfone membranes used several years ago (8,9,10). Study of albumin loss with various combinations of dialyzers and reprocessing techniques is ongoing in the HEMO Study. Preliminary results suggest that the albumin loss from F80B dialyzers currently used in the HEMO Study and reprocessed with bleach is substantially lower than that reported previously.
Pore Theory
The observations in the present study can be explained on the basis of a theory that encompasses two populations of pore sizes on the dialysis membranes. Regardless of the reprocessing technique, reuse of dialyzers often leads to a decrease in effective surface area, thus the number of small pores. This can occur, for example, as a result of clogging of fibers by cells and proteins even though the residual total cell volume remains above 80% of the original value. The loss of small pores then leads to a decrease in urea clearance. In support of this hypothesis, regional loss of flow in the blood compartment of reused dialyzers, suggesting the obstruction of fibers, has been recently demonstrated by a magnetic resonance velocity imaging technique (38). The structure of the residual membrane surface area may also be altered as the result of reuse, the nature and extent of which depend on the reprocessing technique and the membrane material. The size of the large pores may diminish as a result of protein layering, thus decreasing β2M transport, as in the case of cellulose triacetate membrane reprocessed with Renalin. Alternatively, the pores may enlarge, thus enhancing β2M transport, as in the case of polysulfone membrane treated with either heated citric acid or bleach in conjunction with formaldehyde.
Conclusion
To the extent that enhanced removal of small solutes and perhaps middle molecules improves patient outcome, detailed knowledge of the effects of dialyzer reuse on performance is important. The present study illustrates that such knowledge cannot be attained without precise information about the dialyzer membrane structure and the reprocessing technique.
Acknowledgments
Acknowledgments
The study coordinators of the 15 HEMO clinical centers were helpful in data collection. Guofen Yan assisted in data analyses, and Brooks Rogers provided valuable advice regarding reuse. Helpful discussions with Jay Radovich are appreciated.
Footnotes
Dr. Richard A. Ward served as Guest Editor and supervised the review and final disposition of this manuscript.
American Society of Nephrology
- © 1999 American Society of Nephrology